Sensitive Electrochemical Determination of Trifluralin at a Disposable Pencil Graphite Electrode


Gege U., Karakaya S., Dilgin Y.

ELECTROANALYSIS, cilt.33, sa.6, ss.1547-1556, 2021 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 33 Sayı: 6
  • Basım Tarihi: 2021
  • Doi Numarası: 10.1002/elan.202060618
  • Dergi Adı: ELECTROANALYSIS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, Aerospace Database, Applied Science & Technology Source, BIOSIS, Chemical Abstracts Core, Chimica, Communication Abstracts, Metadex, Civil Engineering Abstracts
  • Sayfa Sayıları: ss.1547-1556
  • Anahtar Kelimeler: Differential Pulse Voltammetry, Electrochemical Sensor, Pencil Graphite Electrode, Pesticide, Trifluralin
  • Çanakkale Onsekiz Mart Üniversitesi Adresli: Evet

Özet

A sensitive differential pulse (DP) voltammetric method has been proposed for the determination of trifluralin (TFA) based on both its reduction and oxidation at a disposable pencil graphite electrode (PGE). DP voltammograms recorded under optimized conditions show that oxidation and reduction peak currents increased linearly in the range from 1.0 to 75.0 mu M and from 0.50 to 100.0 mu M TFA, respectively. LOD and sensitivity values have been determined as 0.39 mu M and 11170 mu A mM(-1) cm(-2) for oxidation and as 0.20 mu M and 22167 mu A mM(-1) cm(-2) for reduction. The acceptable recovery values (95.2-104.8 %) were obtained from real water samples.